Products

MODEL 214 STRAIN GAGE BRIDGE CALIBRATOR

Description

The Model 214 provides AC calibration to a strain gage bridge. It will switch a parallel CAL resistor across any leg of the bridge at the selected frequency. The desired CAL resistor is selected using the front panel decade resistance thumbwheel. Any resistance from zero ohms to 9.999 Megohms is selectable in steps of
1K ohm. A front panel LED is provided as a power
on indicator and a low battery indicator. If the battery voltage falls below 15.0 volts the front panel LED will
go out. The batteries are located in a rear panel compartment for easy access.
   


Specifications

Bridge: 120 ohm to 1K ohm


Bridge excitation: 0.5V to 10V, +P to - P


Frequency: Switch selectable at 10, 20, 50, 100,
200, 500, 1kHz


CAL resistor: Adjustable from zero ohms to 9.999 Megohms in 1K ohm steps with ± 1% accuracy


Output: FET switch closure in series with CAL
resistor


Indicators: Combined front panel Power On/Low
Battery


Power: Operates on two 9V batteries, Eveready #505
or equivalent; eight-hour continuous operation


Operating temperature: -25° to 140°F


Rear panel connectors: Monitor banana jacks;
output BNC


Mechanical: High impact plastic instrument case
4.5”W x 8.75”D x 3.5”H

 